Type 202565
JUMO AQUIS 500 CR
The display JUMO AQUIS 500 shows 8888. What is the reason?
- Temperature sensor circuit break
- Probe short circuit on conductivity/temperature sensor
- When using a two-wire measuring cell, there are no bridges in terminals 1 and 2 or 3 and 4 of terminal strip 2 - bridges are included in the package.
- No temperature sensor is connected or connected incorrectly.
- The measured conductivity is higher than the set measuring range by JUMO AQUIS 500.
What applications is JUMO AQUIS 500 CR intended for?
Is a demo version of the JUMO AQUIS 500 setup program available?
30-day trial version is available to test the installation software.
This means that all features of the installation software can be used without restrictions for 30 days after the first installation. Transfer of setting data and saving of setting file are automatically blocked after 30 days.
By entering the license number for the full version, the installer's features can be unlocked again without any time limit (even after 30 days).
The software can be downloaded from the JUMO website at
Why are all the fields in the setup program gray and unavailable for entry?
Can the device be retrofitted with relay or analogue outputs on site?
After pressing CAL a padlock appears next to the calibration type. Why is calibration not available?
What does the value TDS mean?
The TDS (Total Dissolved Solids) value indicates the amount of dissolved salts (including solids) in the water. Common units are ppm or mg/l.
The ppm value is calculated as follows:
ppm = coefficient * value in µS/cm
The coefficient has a default average => 0,67.
What is the difference between two-electrode and four-electrode measuring systems?
What are the advantages and limitations of conductometric conductivity measurement?
Benefits:
- Measuring range depending on the cell constant from 0,05 µS/cm to 200 mS/cm
- Low cost measurement method
- Small design type
Disadvantages:
- Relatively susceptible to contamination
- Measurement error due to line resistance from approximately 50 mS
On what principle is conductometric conductivity measurement based?
The applied voltage is an alternating voltage.
Direct voltage cannot be used because direct current is inevitably associated with decomposition phenomena (polarization phenomena).
The current that is generated is a direct measure of the conductivity prevailing in the medium.
How to restore the device to factory settings?
Then JUMO AQUIS 500 performs a reset. This process takes approximately 10 seconds.
Be careful!
Therefore, the function should only be used if a complete application-specific configuration is available, or if all transmitter requirements for the application are known in detail. Presets from third parties (such as system manufacturers) are reset to standard JUMO and thus lost!
How to find out the software version JUMO AQUIS 500?
Where do I switch the units JUMO AQUIS 500 from mS to µS?
This parameter can be entered in the "Basic Settings" section.
Here you can define various settings:
- Cell type
- Cell constant
- Probe break detection
- Operating mode
- Temperature compensation
- Temperature coefficient
- Unit of measurement
- Decimal place
- Automatic range
After changing the operating mode, the device asks whether it should be restarted.
If there is no response, the old settings are saved. After initialization, the basic settings corresponding to the selected operating mode are applied.