Is JUMO CTI-500 electrically isolated?
Yes, the measuring cell and analog outputs as well as the voltage source and switching outputs are electrically isolated.
What is the CIP-process (Clean in Place / Process)?
CIP is a concept in which the cleaning and disinfection solution circulates in the circuit without dismantling the system to be treated. This is called circulation, a cleaning recirculation circuit or process, which is mainly controlled automatically by programs.
The cleaning concept CIP can be used in systems where circuits can be installed, for example in tank systems, processing lines for drinking milk production, butter production, cheese production, milk powder production or the beverage industry (eg breweries).
The purpose of conductivity measurement technology is to monitor the concentration of cleaning and disinfection solutions in supply containers and at critical points in the process. The concentration in the container for linden and acids must be constant due to intensification with concentrated linden or acid. The conductivity of hot water must be controlled since it is the liquid that comes into contact with the product.
What is the function of desalination or biocide addition?
The desalination function is an automatic process in which the reference value is reduced by opening the desalination valve after a set limit value is exceeded.
The water supply that is now started (control level) dilutes the water in the circuit and thus reduces its conductivity.
In cooling towers, for example, biocides are added to the cooling water to prevent the growth of biological matter in the cooling systems. The desalination lock is activated after the dosing time to ensure that the biocide does not enter the sewer system.
Which measuring cell and which transmitter of the split version belong to each other?
Make sure that 19-digit F No. on the transmitter number plate corresponds to 19-digit F No. at the end of the measuring cable.
If the numbers do not match, calibrate the transmitter and measuring cell using the calibration kit (sales number 20/00459436) (see 3 in FAQ).
What are the repair options?

For head transmitter:
Repairs are only possible at the factory.

With a separate cell:
The measuring probe can be replaced. After replacement, calibrate the probe using the calibration kit (sales number 20/00459436) (see item 3 in FAQ).

What is the correct position for JUMO CTI-500?
  • JUMO CTI-500 can be installed in any position.
  • Avoid direct sunlight.
  • When installing in a pipe system, ensure a minimum distance of 20 mm between the probe and the pipe wall to achieve good flow through and around the probe.
  • If the minimum distance cannot be maintained, the Fitting Factor parameter allows limited compensation.
How to calibrate a measuring cell for a transmitter?
Measuring cell calibration is only required after replacing the measuring cell or transmitter.
For the procedure, please refer to the calibration instructions.
How can I change my working language?

The language can be changed using the setup program or directly on the device.
The device software 196.02.01 provides eight device languages: German, English, Spanish, French, Polish, Swedish, Italian and Portuguese.
Older devices may not be updated in all modern languages.

Changing the language using the installer:

  1. Open the installer
  2. Connect your device to settings
  3. Read data from device
  4. Select device details
  5. Change language
  6. Transferring data back to the device

Changing the language on the device:
The device language is changed at the operator level.

  1. Press the PGM key for more than 3 seconds.
  2. Activate the device at administrator level (standard password 300 can be changed by the customer)
  3. Use PGM-key to select parameter level
  4. Use the arrow key to navigate to the device data option and select with the PGM key
  5. Use the PGM-key to select the language option
  6. Select the desired language using the arrow keys and confirm with the key PGM
  7. Press the EXIT key to exit the parameter level.
Where should the split version be used?
In pipe systems subject to vibrations and high temperatures.