Logo JUMO

JUMO dTRANS Rw 01

Type 202545

Table of Contents

What must be considered when measuring ultrapure water?

Measurement and calibration must be carried out in a closed measuring loop or in a closed container.
Any contact of the medium with the atmosphere leads to a noticeable change in its conductivity.
For this purpose, we recommend using a measuring cell with inspection certificate ASTM.

The device displays an error message “ERR”.

Calibration of the relative cell constant or determination of the temperature coefficient of the measured medium resulted in an error. Previous data is retained.
Factory data is displayed (Cell = 100%, ALPH = 2.3).
To correct the "ERR" error message, manually change two parameters using the keypad (for example, change one digit up or down and confirm with the PGM key) or recalibrate the system.

Actual performance required - what should you consider?

Make sure your device has the appropriate analog output. For this purpose, check the device type designation.
See chapter 4.1, Explanation of type, in the operating instructions.

​
​
Configuration level C213 - other I outputs are installed.
The actual output value 2 is set in configuration level C214 - Other outputs II.
Scaling is set using parameters SOL and SOH.

I can't change the parameter level and configuration level settings.

To be able to make settings on two levels, first enter the corresponding CODE to activate.
Start Situation > Measurement Mode > Press the PGM key repeatedly until CODE appears. Set the corresponding CODE by pressing the UP/DOWN keys. Press key PGM to confirm.
The display returns to 0000. Now you can select the appropriate level (parameter). The code is entered correctly when the decimal point flashes during entry.

Function

Code

Unlock operator level, CAL and manual Hold activation.

0110

Unlock operator and parameter level.

0220

Unlock all levels.

0300

Is it possible to change the N/C/N/O switching behavior?

The N/O/N/C setting for the two pins K1 and K2 can be changed at the configuration level under point C212 (controller outputs).
Both contacts are factory set NOC.
NOC: (N/O) The corresponding output is active (closed) as long as the switching condition is met.
NCC: (N/C) The corresponding output is active (closed) until the switching condition is met.

What does pin MIN/Max mean?

Configuration level C212 (controller pins) can be used to change the Min/Max behavior for the two pins K1 and K2.
MIN-contact: The switching condition is fulfilled when the actual value is lower than the set point value.
Contact MAX: The switching condition is satisfied when the actual value exceeds the set point value.

​
​

How to check the correct function (actual display value) of the device and coaxial cable?

The Redox value can also be set using the pH simulator as for pH measurement. The Redox value can then be read from a table contained in the simulator's operating instructions. If a simulator is not available, voltage can be applied directly to the input of the device (receptacle BNC) and, for testing coaxial cable, to the end of the cable.
The Rd value can be read on the device according to the voltage.
The device should be in factory settings and should not display any error messages.

pH value

Voltage

2 pH

+295,8 mV

4 pH

+177,48 mV

7 pH

0 mV

10 pH

-177,48 mV

10 pH

-295,8 mV

​
​

How to start the calibration process?

To begin calibration, activate the controller.
Start situation > Measurement mode > Press the PGM key repeatedly until CODE appears > Enter 0300 and confirm with PGM > Press the Exit key once > Press the PGM + Down key at the same time.
Temperature 25,0 appears in the upper (red) display, °C appears in the lower (green) display, the decimal point in the lower display flashes.

Our services for you

Privacy Settings

This preview uses local files. Analytics services JUMO are not connected.